DNA Sequencing and Fragment Analysis

Posts tagged ‘Template Preparation’

Protocol: Alternative Method for Sodium Acetate to Ethanol Precipitate DNA

Ethanol with sodium acetate is used as a method for precipitating DNA in solution. It is preferred over simple drying in order to concentrate DNA. Drying concentrates salts and contaminants in addition to the DNA. Ethanol precipitation allows contaminants to be removed along with the excess liquid while the DNA forms a solid pellet on the bottom of the tube. It is a good method. Isopropanol is also used as a replacement for ethanol. It requires less volume. However, we recommend ethanol, as it seems to clean the DNA more efficiently.

Sodium acetate is a salt. The salt is a necessary ingredient because it acts as a carrier for the DNA. Without sodium acetate, the majority of the DNA would remain in solution. It would be lost along with the other contaminants.

The following method works well for final purification of PCR fragments and plasmid preparations with a volume of 20 to 25 ul. Glycogen replaces sodium acetate as the DNA carrier.

Things You Need

95 % Ethanol

70 % Ethanol

Glycogen (20 mg/ ml)

Microcentrifuge tubes (1.5 ml)

Microcentrifuge

Hot plate with tube block

Protocol

Step 1: Add 1.5 ul glycogen to DNA in solution. It is important to add the glycogen before adding ethanol because ethanol will cause glycogen to precipitate.

Step 2: Add 80 ul 95 % ethanol.

Step 3: Vortex to mix components.

Step 4: Allow DNA and glycogen to precipitate for a minimum of 15 minutes.

Step 5: Centrifuge for 15 minutes to pellet DNA.

Step 6: Remove excess ethanol by decanting or using a pipette.

Step 7: Add 200 ul 70 % ethanol

Step 8: Mix gently by inverting the tube several times.

Step 9: Centrifuge for 15 minutes.

Step 10: Remove excess ethanol by decanting or using a pipette.

Step 11: Dry in heat block to remove remaining ethanol.

Step 12: Re-suspend in 10 ul purified water.

Step 13: Determine concentration.

Variations to the Protocol

Temperature and time are 2 variables that effect the precipitation of DNA. Often precipitation is performed at 4O C to improve yields. Time is also a factor. Precipitating for longer periods could improve yields. However, using lower temperatures and longer precipitation times could drop certain contaminants out of solution. Some caution should be used when changing both variables.

Once the DNA pellet is dry, it could be re-suspended in water or a buffer of choice. Water has the advantage of adding no additional salts initially removed during the precipitation. The disadvantage of water is that DNA is more easily subject to degradation. Water is a reasonable diluent for short-term applications or Sanger sequencing. Some sequencing facilities recommend water for the final aliquot in preparation of samples. TRIS and TRIS with EDTA are also used for long-term storage of DNA samples. Buffers containing EDTA are not recommended for samples submitted for Sanger sequencing and PCR because EDTA interferes with the MgCl2, a necessary component for amplification.

Please submit any additional suggestions or questions related to the glycogen protocol for ethanol precipitation of DNA.

Please go here if you would like to download a

reprint for this article in pdf format

Double Sequence: A Common Problem in Sanger Sequencing

Sanger sequencing methodology requires template DNA to be relatively free from contaminating salts. Template quality is often determined electrophotometrically by loading the template on agarose gel. Agarose gel electrophoresis often shows the faster moving supercoiled DNA and slower moving nicked DNA found in a quality preparation. DNA samples produced by PCR amplification can be similarly tested by agarose gel electrophoresis. A good PCR product is viewed as a single clean band representing the desired length.

Another test is to measure absorbance spectrophotometrically at ultraviolet wavelengths 260 nm and 280 nm. A 260 nm / 280 nm ratio value of 1.8 is an indicator that the DNA is good.

Unfortunately, the result off the capillary sequencer shows double set of peaks. Because one set of peaks is directly over top the other, software cannot determine the correct bases to call as in figure 1. A result showing a double set of peaks is not as uncommon as many might believe.

What are the primary causes of double sequence?

Double Sequence Caused by Template Contamination

One of the early steps in plasmid DNA preparation is to select a desired colony and then inoculate liquid medium. In an overnight culture, the inoculated media would go through rapid bacterial cell division as the cells enter the log growth phase. Because the initial cells came from a single colony, it is believed that every cell is the identical product cloned from a single cell. However, this is not always the case. Colonies do not always come from a single cell. A colony could also be the product of multiple cells that failed to separate when spread on a plate. If the multiple cells are not identical, the sequence result could produce the double sequence shown in figure 1.

The region where double sequence begins is usually a restriction sight where a separate piece of DNA is inserted into a vector. One purpose of bacterial cloning is to produce copies of the inserted DNA. The clean sequence shown in figure 1 is vector sequence before the restriction sight. Despite differences in the two templates, the vector sequence is generally the same. The template contamination becomes apparent at the point where the new DNA has been inserted into the vector.

Often, the reverse or complimentary strand does not produce a double sequence and this is often confusing. If the DNA insert were relatively long, it would require sequencing the whole complimentary strand before double sequence is observed. The reason is simply because double sequence is caused by addition or deletion of an unknown number of bases at one restriction sight only. Therefore, both templates, the desired and contaminant have identical sequence in the complimentary direction.

PCR fragments could also produce double sequence. It is the result of errors in PCR amplification where two products are produced. The PCR fragments could produce two very similar sized products that appear as a single band once tested by agarose gel electrophoresis. Resolution using agarose gel is limited and does not separate DNA samples of very similar size.

Double Priming Results in Dual Amplification

Sanger sequencing is similar to PCR amplification. A primer anneals at the beginning of the region to be sequenced and Taq polymerase adds bases (dNTPs) in extension to produce and identical strand. A primer that matches two regions on either the insert or the plasmid could cause two separate amplifications simultaneously. The result of double priming is shown in figure 2.

Unlike template contamination, two separate products generally do not appear clean in the beginning. It does not require the primer to match identically to both annealing sights as long as the bases of the primer are similar with a match on the 3 prime ends. What happens if one annealing site is downstream from the other annealing site on the same strand of DNA? The downstream primer blocks bases extending from the first primer and the double sequence eventually becomes a single set of peaks.

Single Nucleotide Polymorphisms Could Also Cause Double Sequence

PCR amplification is used to determine potential heterozygous bases as single nucleotide polymorphisms (SNPs). It is one method for detecting mutations that could cause certain genetic diseases. The SNP likely could be observed as a base change in one of two identical alleles amplified together and is generally easy to spot from automated Sanger sequencing. A single base position is represented by two peaks to positively identify a SNP.

However, SNPs could also be the result of a base insertion or deletion (indel) as in figure 3. From the figure it is possible to follow two identical strands of sequence that are different by one base. Both alleles match until the GGCC region. However, one allele is missing the T-base and has shifted upstream by one base. The large clean peaks observed in the double peak region are simply the same base represented by two different products.

Preventing Double Sequencing in Sequencing Results

There are some cases where double sequence cannot be resolved as with PCR fragments that contain indels. However, certain steps could be taken to reduce conditions described in figure 1 and figure 2.

Plasmid preparations that produce double sequence at the point of insert can be prevented using a process called single colony isolation. Once the plasmid cells have been spread on a plate and grown, the select colony is isolated and spread on a second plate. Although the process requires additional time, it is worth the effort in cases when double sequence often occurs.

The process for preventing double sequence in PCR fragments is more complicated because PCR conditions need better optimization. It could be something as simple as reducing the amount of primer or number of cycles. Or require small adjustments in PCR conditions including temperature. Guidelines for PCR optimization will be a future topic.

Samples that produce double sequence as the result of double priming can be resolved by extending bases to the custom primer that was selected for sequencing. A better choice is to select a different primer even when double priming is caused from one of the possible universal primers present on a particular vector.

Double sequence result patterns are generally similar for each condition discussed here. Anyone is invited to present results similar to that presented. Questions are also welcome.

Please go here if you would like to download a

reprint for this article in pdf format

How to Determine if a Sequencing Template Meets Quality Requirements

Isolation of plasmid DNA in preparation for automated Sanger sequencing has been simplified by the development of commercial kits. The quality of the final sample is generally tested by spectrophotometer measuring absorbance at wavelengths 260 nm and 280 nm. A ratio value of 1.8 calculated for absorbance measurements at 260 nm by 280 nm indicates the DNA sample is good. However, 260 nm and 280 nm absorbance values do not provide a complete profile of whether DNA preparations meet necessary quality standards for Sanger sequencing applications. Samples may still fail. There are a number of reasons why sequencing reactions are not successful. Two well-known techniques eliminate template quality as a reason for failure.

Scanning Spectrophotometry Provides Important Information about Template Quality

Nanodrop technology has the advantage of highly sensitive spectrophotometric scanning of DNA samples using just 1 ul of volume. Most spectrophotometers available today are capable of performing a wavelength absorbance scan in the ultra violet spectrum. DNA is scanned between wavelengths ranging from 220 nm to 320 nm. The scan profile for quality DNA is shown in figure 1.

A typical DNA profile is a Gaussian (bell-shaped) curve with the maximum peak height absorbing at 260 nm. A secondary peak will also begin to form around 220 nm. Deviations in the shape of either curve could indicate the presence of salts or other contaminants.

Figure 2 provides a profile of plasmid DNA containing salts compared to clean DNA. The Gaussian curve should drop almost to the baseline when DNA is free of contaminating salts in the range 230 nm to 240 nm. Salts present in the sample will absorb wavelengths in this range. Excessive amounts of salt in DNA samples may show little or no dip in the curve between 220 nm and 260 nm.

Figure 3 represents a profile of DNA that contains either protein (DNA from tissue) or phenol (plasmid DNA). Phenol residue in DNA can result from phenol/ chloroform extraction (a method of extraction that was widely used prior to the development of commercial kits). Despite the availability of kit-based methods, many researchers opt to use phenol/ chloroform instead. It is important to remove phenol residues with ethanol precipitation in the final purification step. Phenol residue can negatively affect the PCR amplification in Sanger sequencing.

Scanning spectrophotometry successfully identifies salt contaminants in plasmid DNA and DNA samples amplified in the Polymerase Chain Reaction (PCR). PCR fragments will have a lower concentration. But purified PCR products should still show a scan similar to plasmid preparations.

Agarose Gel Electrophoresis Identifies Nicked DNA

Once plasmid DNA has been determined to be free from salt contaminants it should work well for Sanger sequencing. However, this is not always the end result. Scanning spectrophotometry does not reveal whether DNA is supercoiled or nicked as described previously (Nicked Plasmid DNA Prevents Automated Sanger Sequencing). Nicked DNA has been damaged mechanically. Aggressive vortexing during DNA purification is one cause of DNA damage. Nicked DNA does not amplify in Sanger sequencing applications because the double stranded helix does not maintain a tight formation. Taq polymerase is unable to fulfill a lock-key attachment to the DNA and catalyze extension. For automated sequencing to work, plasmid DNA must maintain a supercoiled structure.

Agarose gel electrophoresis successfully separates nicked DNA (slow migrating) and supercoiled (fast migrating) as shown in figure 4. Nicked DNA moves more slowly through the gel because it is a larger molecule than supercoiled DNA. It should be noted that plasmid preparations typically have a mixture of nicked and supercoiled DNA. Presence of some nicked DNA should not cause concern. It is the lack of supercoiled DNA that causes sequencing failures.

Scanning spectrophotometry and agarose gel electrophoresis combined provide a good assessment of whether template DNA purification has been successful. However, it does not guarantee Sanger sequencing will be successful every time. Primer selection, GC content and presence of poly base regions and hairpins could also affect sequencing. Problems of this nature are related more to base content and not preparation of the sample.

Please go here if you would like to download a

reprint for this article in pdf format

Preparing Plasmid DNA for Automated Sanger Sequencing

Cesium chloride and phenyl chloroform were two early methods used for isolating a sample of DNA from a bacterial vector. Cesium chloride produced a sample that was very clean. However, the method was labor intensive. Phenyl chloroform was less labor intensive, but phenyl contamination was difficult to remove even using the final alcohol precipitation step. Fortunately, commercial kits were developed around the same time Sanger sequencing became automated. Presently there are numerous commercial kits from which a researcher could choose.

However, using a commercial kit does not guarantee that a resulting DNA sample will be clean and free from contaminants. The final step of the sample preparation often includes Tris or another buffer (salt) for elution. Other times the DNA may incur physical damage and become nicked. Preparing a sample for automated sequencing includes following the recommended guidelines provided for the commercial kit. However, low yields can require additional steps to concentrate the sample. Simple drying methods concentrate everything in the sample including buffer salts.

Steps Involved in Commercial DNA Preparation Kits

The general protocol for commercial kits used to isolate purified DNA from plasmids is fairly similar regardless of the kit. After plasmid bacteria have been grown overnight in liquid media, the cells are pelleted by centrifugation into a solid mass. The media is then discarded before beginning the kit provided procedure (Figure 1).

The pellet of cells is resuspended in buffer and transferred into a 1.5 or 2 ml tube (Step 1). Lysis reagent is added (Step 2). The lysis reagent disrupts the bacterial cell membranes freeing internal components. The lysis buffer is neutralized with neutralizing solution and cell components form a precipitate with the DNA still in solution (Step 3). The resulting solution is centrifuged to pellet the cell waste leaving the DNA in solution.

DNA is captured on a filter or resin by transferring the DNA solution to a collection vessel (Step 4). Excess liquid is pulled through the filter by centrifugation or vacuum filtration. The remaining DNA on the filter is then washed removing any cell waste left (step 5). The DNA is then collected in a clean tube using water or elution buffer (step 6). Elution buffer is provided in the commercial kits. It is comprised of Tris buffer in water. Fortunately, manufacturers no longer include EDTA in elution buffer. EDTA interferes with magnesium chloride, a necessary component in any PCR amplification. Water is the preferred media for DNA to be used for Sanger sequencing as no salts are being introduced to the sample.

The process may take as little as 15 minutes after bacterial cells are grown overnight. The final DNA sample is relatively pure and clean from other cellular debris that could interfere with Sanger sequencing.

Quality Testing of the DNA Sample

DNA sample quality can be determined using two simple methods. The methods will be discussed in detail in the next article, but are summarized below.

Scanning spectrophotometry using ultraviolet wavelengths between 220 nm and 310 nm provides a general profile of the overall quality of the DNA (Figure 2). DNA absorbs light in the range of 240 nm to 300 nm with the maximum peak at 260 nm. Absorbance at 260 nm is also used to calculate the concentration of the DNA. Another factor often used when measuring DNA quality is calculating the 260 nm / 280 nm absorbance ratio. A good value for this ratio is 1.8 to 1.9. Ratio values below 1.6 indicate the DNA sample contains contaminants.

A good scanning profile does not always prove the DNA is good. Nicked DNA cannot be identified by scanning and may inhibit quality Sanger sequencing. (http://agctsequencing.wordpress.com/2012/02/16/nicked-plasmid-dna-prevents-automated-sanger-sequencing/)

Agarose gel electrophoresis will separate DNA into bands indicating whether the DNA has remained supercoiled.  This supercoiled DNA is required for sequencing. It will also show if the DNA has been damaged and the supercoil has loosened when the DNA is nicked. Nicked DNA migrates more slowly through an agarose gel and will separate from the supercoiled DNA.

Both quality tests provide necessary data to show whether the final DNA sample is clean and of high quality.

Additional Concentration Required

Once the DNA sample has been isolated and appears clean using test procedures, it may be necessary to adjust the concentration to meet submission guidelines. Samples with a concentration higher than required are diluted in water. Although some buffer is still present, the buffer is diluted along with the DNA. Generally dilute buffers (without EDTA) do not interfere with amplification for Sanger sequencing. What if the sample needs to be more concentrated?

Drying a sample in elution buffer is not an effective means for concentrating a sample. Ethanol precipitation is a preferred method because salts are removed along with excess water. Scanning spectrophotometry is an effective means of quantifying the DNA.

What is most important for researchers to remember when isolating plasmid DNA with a commercial kit is to thoroughly read the directions. The kits use enzymes to disrupt the bacterial cell membrane and remove components such as RNA. Enzymes are fragile. Excess shaking or vortexing could damage the enzymes. Commercial kits often include precautions to use care when working with these important enzymes.

Please go here if you would like to download a

reprint for this article in pdf format