DNA Sequencing and Fragment Analysis

Cesium chloride and phenyl chloroform were two early methods used for isolating a sample of DNA from a bacterial vector. Cesium chloride produced a sample that was very clean. However, the method was labor intensive. Phenyl chloroform was less labor intensive, but phenyl contamination was difficult to remove even using the final alcohol precipitation step. Fortunately, commercial kits were developed around the same time Sanger sequencing became automated. Presently there are numerous commercial kits from which a researcher could choose.

However, using a commercial kit does not guarantee that a resulting DNA sample will be clean and free from contaminants. The final step of the sample preparation often includes Tris or another buffer (salt) for elution. Other times the DNA may incur physical damage and become nicked. Preparing a sample for automated sequencing includes following the recommended guidelines provided for the commercial kit. However, low yields can require additional steps to concentrate the sample. Simple drying methods concentrate everything in the sample including buffer salts.

Steps Involved in Commercial DNA Preparation Kits

The general protocol for commercial kits used to isolate purified DNA from plasmids is fairly similar regardless of the kit. After plasmid bacteria have been grown overnight in liquid media, the cells are pelleted by centrifugation into a solid mass. The media is then discarded before beginning the kit provided procedure (Figure 1).

The pellet of cells is resuspended in buffer and transferred into a 1.5 or 2 ml tube (Step 1). Lysis reagent is added (Step 2). The lysis reagent disrupts the bacterial cell membranes freeing internal components. The lysis buffer is neutralized with neutralizing solution and cell components form a precipitate with the DNA still in solution (Step 3). The resulting solution is centrifuged to pellet the cell waste leaving the DNA in solution.

DNA is captured on a filter or resin by transferring the DNA solution to a collection vessel (Step 4). Excess liquid is pulled through the filter by centrifugation or vacuum filtration. The remaining DNA on the filter is then washed removing any cell waste left (step 5). The DNA is then collected in a clean tube using water or elution buffer (step 6). Elution buffer is provided in the commercial kits. It is comprised of Tris buffer in water. Fortunately, manufacturers no longer include EDTA in elution buffer. EDTA interferes with magnesium chloride, a necessary component in any PCR amplification. Water is the preferred media for DNA to be used for Sanger sequencing as no salts are being introduced to the sample.

The process may take as little as 15 minutes after bacterial cells are grown overnight. The final DNA sample is relatively pure and clean from other cellular debris that could interfere with Sanger sequencing.

Quality Testing of the DNA Sample

DNA sample quality can be determined using two simple methods. The methods will be discussed in detail in the next article, but are summarized below.

Scanning spectrophotometry using ultraviolet wavelengths between 220 nm and 310 nm provides a general profile of the overall quality of the DNA (Figure 2). DNA absorbs light in the range of 240 nm to 300 nm with the maximum peak at 260 nm. Absorbance at 260 nm is also used to calculate the concentration of the DNA. Another factor often used when measuring DNA quality is calculating the 260 nm / 280 nm absorbance ratio. A good value for this ratio is 1.8 to 1.9. Ratio values below 1.6 indicate the DNA sample contains contaminants.

A good scanning profile does not always prove the DNA is good. Nicked DNA cannot be identified by scanning and may inhibit quality Sanger sequencing. (https://agctsequencing.wordpress.com/2012/02/16/nicked-plasmid-dna-prevents-automated-sanger-sequencing/)

Agarose gel electrophoresis will separate DNA into bands indicating whether the DNA has remained supercoiled.  This supercoiled DNA is required for sequencing. It will also show if the DNA has been damaged and the supercoil has loosened when the DNA is nicked. Nicked DNA migrates more slowly through an agarose gel and will separate from the supercoiled DNA.

Both quality tests provide necessary data to show whether the final DNA sample is clean and of high quality.

Additional Concentration Required

Once the DNA sample has been isolated and appears clean using test procedures, it may be necessary to adjust the concentration to meet submission guidelines. Samples with a concentration higher than required are diluted in water. Although some buffer is still present, the buffer is diluted along with the DNA. Generally dilute buffers (without EDTA) do not interfere with amplification for Sanger sequencing. What if the sample needs to be more concentrated?

Drying a sample in elution buffer is not an effective means for concentrating a sample. Ethanol precipitation is a preferred method because salts are removed along with excess water. Scanning spectrophotometry is an effective means of quantifying the DNA.

What is most important for researchers to remember when isolating plasmid DNA with a commercial kit is to thoroughly read the directions. The kits use enzymes to disrupt the bacterial cell membrane and remove components such as RNA. Enzymes are fragile. Excess shaking or vortexing could damage the enzymes. Commercial kits often include precautions to use care when working with these important enzymes.

Please go here if you would like to download a

reprint for this article in pdf format

Advertisement

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out /  Change )

Facebook photo

You are commenting using your Facebook account. Log Out /  Change )

Connecting to %s

%d bloggers like this: